SUPPLEMENT MOVING FRONT CLASS 5
TYPE 5 MOVING FRONT SUPPLEMENT

- pack of 1,000 pcs
- non-toxic/lead-free

Moving Front 134°C -3.5 min integrators give the best possible interpretation in case of any failure in steam sterilisation cycles.
If the temperature is not reached or the holding time is too short, the dark colour will not reach the acceptance area.
Conforms to ISO 11140 - type 5.

User manual in GB, IT, FR, ES.

- shelf life: 5 years from date of manufacture
- white aluminium packaging, re-sealable bag
- easy-to-interpret result
- sealed indication area
BOWIE DICK PACK
BOWIE AND DICK TEST PACKAGE

- pack of 20 pcs
- dimensions: 123 x 123 x 16 mm

Used for daily monitoring of pre-vacuum steam sterilisers.
The Bowie and Dick test is intended for testing the steam steriliser at the beginning of the day (as per ISO 11140).
If the test is successful, it authorises the use of the steriliser.
The Bowie and Dick test detects faults such as air leaks, insufficient vacuum and inadequate steam penetration.
Packaging and manual in GB, IT, ES, FR, PT, GR, DE.
- meets EN 285 and ISO 17665 requirements
- complies with ISO 11140 - 4 TYPE 2 / EN 867 - 4 CLASS B (EU standard 7 kilo)
- independently tested to show similar performance to the original package
- light blue - violet colour change
- easy to interpret and document
- non-toxic/lead-free
ETHYLENE OXIDE TEST - double
ETHYLENE OXIDE STERILIZER INDICATOR - pack of 250 double strips

Ethylene Oxide (EO) type 5 steriliser indicators are designed for use with E.O. sterilisers in accordance with ISO 11135.
The indicator strips can be used in any instrument pouch and provide guaranteed penetration of E.O. gas into the instrument pouch.
The water-based, non-chemically toxic indicator in accordance with ISO 11140 - 1 will show a clear colour change from white to yellow after sterilisation.
The strip test indicators are convenient and inexpensive and can be split into two separate pieces.
STERILISATION TEST CLASS 6
EMULATION INDICATOR (STEAM) TYPE 6 - pack of 250 strips

Emulation indicators (type 6) are designed to test steam sterilisers operating at 121°C - 15 min and 134°C - 3.5 min.
The test indicators can be used in any bag/pack of instruments and will provide assurance of steam penetration into the instruments and the presence of all critical parameters required in the steam sterilisation cycle.
The water-based, non-chemically toxic indicator in accordance with ISO 11140 - 1 will show a clear blue to lilac colour change after sterilisation.
SEAL CHECHER - tests for sealers
SEALING CONTROL

- pack of 250 pcs
- dimensions: 175 x 75
- suitable for all rotation sealers

The ISO 11607 standard specifies a daily check routine for heatseals.
The seal check is a method of performing the required test, documenting the day, the person who performed the test and clearly showing the results of your traysealer's ability to properly heat and seal the bag.
When it is placed inside a subsequently sealed bag and the test area becomes darker, it will be easy to interpret whether the seal is compliant, smooth and completely leak-free.
STERILISATION TEST CLASS 4 - double
STEAM STERILIZATION INDICATOR TYPE 4 - pack of 250 double strips

Multi-variable Class 4 steam sterilisation indicators are designed for use in steam sterilisers operating at 134° C - 3.5 min.
The indicator strip can be used in any pouch and will provide guaranteed steam penetration into the instrument packs placed in the steriliser.
The water-based, non-chemically toxic indicator in accordance with ISO 11140 - 1 will show a clear colour change from blue to black after successful steam sterilisation.
The strip indicators are convenient and inexpensive and can be split into two separate pieces.
LEAD-FREE INDICATOR TAPE 3M COMPLY - 1322
3M™ COMPLY™ INDICATOR TAPE WITHOUT LEAD 1322 - 18 mm x 55 m - pack of 28 rolls

Consisting of adhesive, backing and chemical indicator ink strip.

Strong, pressure sensitive adhesive designed to adhere to a variety of wrappings including raw fabric, disposable non-woven, paper and plastic to seal packaging during steam sterilisation.

The beige disposable crepe paper backing allows the necessary elongation for packaging expansion.

The striped chemical indicator ink shows a colour change from whitish/beige to dark brown/black when exposed to sterilisation.

- natural rubber latex-free
- reliable high quality adhesive
- lead-free ink avoids hazardous waste disposal problems
- stretchable backing that does not come off during sterilisation
- possibility of pre-printed lettering or labels
STERILISATION TEST CLASS 5
STEAM STERILISATION INDICATOR TYPE 5

- package of 500 pcs.
- non-toxic/lead-free

Allows reconfirmation of sterilisation conditions inside trays or containers.
Steam Integrator: 134 °C - 3.5 min, 121 °C - 15 min.
Complies with ISO 11140 - type 5.

Packaging and user manual in GB, IT, FR, ES, PT, GR, DE.

- shelf life: 5 years from date of manufacture
- packaged in easy dispensing carton
- clear transition of indicators from blue to pink
MULTIUSI GEL - 150 ml
MULTIUSI GEL

- 150 ml
- pack of 24 pcs

This is a ready-to-use product containing 62% ethyl alcohol for rapid and effective hand hygiene, designed to sanitise and cleanse hands without the use of water.
Its special formulation dries quickly, leaving hands soft and fresh.
Disinfectant and dermoprotective action.
Italian production.
AUTOCLAVE PRESTIGE 12 L with manometer - 12 litres
AUTOCLAVI PRESTIGE - class N

- Autoclave Prestige 12 litres with pressure and temperature gauges

Prestige is one of the cheapest and best-selling automatic autoclaves on the market; its excellent reputation is due to its reliability, durability, safety and quality.

- method recommended by healthcare professionals around the world, including the World Health Organisation (WHO)
- each autoclave is computer tested and issued with an original performance certificate
- lid with pressure and temperature gauge

Recommended for infection control not only in hospitals and medical and dental practices, but also in veterinary practices, podiatry offices, beauty salons, tattoo parlours, acupuncture centres, laboratories and schools.

Complies with IEC 61010 (product safety), BS 3970 part 4 (pressure vessel only), BS 5500 (pressure vessel).

Supplied with basket and multilingual manual: GB, FR, IT, DE, ES, PT, NL, DK, NO, SE, GR. Capacity: 12 l
Internal chamber dimensions (W/H) 210/270
Max. instrument length (diagonal): 268 mm
Maximum load capacity: 4 kg
Width/height: 340/420
Sterilisation time: 11 minutes
Cycle time (nominal): 22 minutes
Working pressure: 1.40